8.3 Calibration
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
87
Figure 8-13 Example AUTOCAL sequence
List for the AUTOCAL sequence:
Note
Zero gas 2 is not used with FIDAMAT
6.
Autocal/-check cyclic parameter
Figure 8-14 AUTOCAL/-check cyclic parameter
With this subfunction, you configure various time constants for activating a cyclically
recurring AUTOCAL.
"Time from autocal to autocal (cycle time)":
Any setting between 1 and 720 (hours) is accepted by the device.
"Time up to next autocal cycle"
(after the time of setting):
If you enter "0" here and AUTOCAL is activated with
"Autocal on/off"
, the device immediately
begins the AUTOCAL sequence.