• When drilling, ensure that the drill is placed vertically against the plate.
• Due to the step length of the drill, a small 26 mm diameter bridge remains on the inner
edge of the hole. Remove this bridge with a half-round file. The diameter drilled should
be 28 mm across the entire plate thickness. Use the file to remove the burr as well.
• Also deburr the two 3.5 mm holes. This can be done with a drill having a diameter of
approx. 8 mm to 10 mm. Place the drill manually against the 3.5 mm holes from the
inside and brake the burr.
• Remove all drill shavings outside and inside with a vacuum cleaner.
• Varnish the 28 mm hole with a varnish applicator and allow it to dry.
• Position the connector with holding plate (1/Fig. 27) from the inside.
• The connector is positioned centrally to the 28 mm hole, the lid opens downward.
• Fasten the connector with holding plate using the screws and lock washers supplied
(1/Fig. 28).