Siemens AG SPR2-230.814.01 Page 1 of 2 SIREMOBIL Iso-C 3D
Medical Solutions Rev. 05 08.04 CS PS 24
4. Part of activities independent of serial number7
Checking the C-arm movements for SIREMOBIL Iso-C 3D 7
Proceed as follows after reconfiguring the basic unit and the monitor trolley.
Connect the cables in the zipper hose to the basic unit.
Connect the monitor trolley to the line voltage.
Release the orbital brake.
Move the C-arm by hand into both end positions. The movement has to correspond to
the "normal" movement of a C-arm. If not, verify that connector M3.X1 for the brake
assembly is inserted correctly. If yes, the LEDs in both keys light up.
Where to go from here 7
Continue with the SIREMOBIL Iso-C 3D Start-up Instructions SPR2-230.815.01... .