7 - 1
Siemens AG SPR2-230.814.01 Page 1 of 2 SIREMOBIL Iso-C 3D
Medical Solutions Rev. 05 08.04 CS PS 24
4. Part of activities independent of serial number7
Checking the C-arm movements for SIREMOBIL Iso-C 3D 7
Proceed as follows after reconfiguring the basic unit and the monitor trolley.
• Connect the cables in the zipper hose to the basic unit.
• Connect the monitor trolley to the line voltage.
• Release the orbital brake.
• Move the C-arm by hand into both end positions. The movement has to correspond to
the "normal" movement of a C-arm. If not, verify that connector M3.X1 for the brake
assembly is inserted correctly. If yes, the LEDs in both keys light up.
Where to go from here 7
Continue with the SIREMOBIL Iso-C 3D Start-up Instructions SPR2-230.815.01... .