SIREMOBIL Iso-C 3D SPR2-230.814.01 Page 2 of 6 Siemens AG
Rev. 05 08.04 CS PS 24 Medical Solutions
1 - 2 General
Safety information 1
Measuring the protective conductor resistance 1
• After all work has been completed and all covers have been attached, perform the
protective conductor measurement according to ARTD-002.731.17.
• The protective conductor resistance must not exceed 0.2 ohms.
When performing the work steps and tests described herein,
observe the product-specific safety information contained in the
documents, as well as the general safety information
TD00-000.860.01....
Prior to performing the work, disconnect the power plug of the
SIREMOBIL Iso-C.
CAUTION
WARNING