3. Part of activities independent of serial number 6 - 41
Siemens AG SPR2-230.814.01 Page 41 of 42 SIREMOBIL Iso-C 3D
Medical Solutions Rev. 05 08.04 CS PS 24
Covers 6
Attaching the PC covers 6
Connect the protective conductor (1000 mm long) from the front cover of the base plate
to the PC support (Fig. 95).
Attach the front cover (Fig. 94).
Connect the protective conductor (150 mm long) from the side cover of the base plate to
the PC support.
Connect the protective conductor (550 mm long) from the rear panel of the PC to the
base plate of the PC support.
Route the protective conductor from the base plate of the PC support to the protective
conductor bar in the monitor trolley and connect it.
Attach the rear panel of the monitor trolley.
Affix the 3D label to the monitor (Fig. 96).
Fig. 95 PC with covers Fig. 96 Grounding of PC housing
Fig. 97 3D label