2-53
1010ENFM-3B
Section 2
Siemens 2 Channel [1] SITE1
Select Data Units or Percent Display
Stripchart Setup
Select Data LiquIdent
Data Display Data Rate Units
Time Base 1 second
Stripchart Clear Yes
Use this menu cell
to select an update
interval for the
stripchart.
Use this menu cell
to clear the current
stripchart.
Use this menu cell
to select a data
item to display on
the stripchart.
Use this menu cell
to select whether
the stripchart
shows data as
either a rate or a
percentage of
span.
Press to move to the [Indicated Rate] field, then to enable numeric entry.
Use the numeric keys to type a flow rate for point #1, then press
NOTE: Enter absolute value of the flow. Flow rate is not direction sensitive. The same calibra-
tion applies to both negative and postitive flow rates.
The highlight moves to the [% Correction] field for point #1. Press to enable numeric
entry.
Use the numeric keys to type the required % correction for point #1.
Press . The highlight returns to the [Point #] field.
To enable numeric entry press then type [2] (for the second calibration point).
Press . The cursor moves to the point #2 [Indicated Rate] field.
Use the numeric keys to type a flow rate for point #2. The value entered here
must be of a greater magnitude than the value entered for point #1.
Press . The cursor moves to the [% Correction] field.
To enable numeric entry press
Use the numeric keys to type the required % correction for point #2. Press
Repeat this procedure for all required calibration points. After entering the last
Flow Calibration Point press to exit MultiPoint calibration.
ENT
ENT
ENT
ENT
ENT
2.8 THE STRIPCHART SETUP MENU
This menu becomes available after picking a Meter Type and Measurement Channel.
Use Stripchart Setup to select a data category, the type of scaling and the update interval for the on-
screen stripchart display. You can also clear the stripchart screen.

Table of Contents