2.11.5 THE APPLICATION INFO MENU
This menu provides a live display of the basic timing data used by meter during operation.
2.11.6 THE LIQUID DATA MENU
This menu shows the current Reynolds number used by the meter to implement flow profile compensa-
tion.
To activate AnCal:
To move the cursor to the [AnCal] menu cell press
Press to enable numeric entry. Note that an equal sign (=) appears before the number.
Type the desired flow rate using current rate units (e.g., 120.00 GPM). Note that the [Flow] menu
cell now reflects the artificial rate.
To move the cursor away from the menu cell press to turn AnCal off.
TN uSec 75.193
TL uSec 44.009
DeltaT uSec 126
Burst/Sec 256
% Accepted 100
Last Makeup (Re)Start
Makeup Status Measurement
Siemens 2 Channel [1] 1
Time between transit and receive
Application Info
APPLICATION INFO MENU ITEMS
TN uSec The total elapsed time between the transmission and reception of a sonic pulse.
TL uSec, The total time a sonic pulse takes to travel through the liquid.
DeltaT nSec This is the instantaneous result of subtracting the down from the up transit-times.
NOTE: This will appear to be more active than dampened flow readings.
Burst/Sec Number of transmissions per second under current operating conditions. The
following factors influence this parameter: aeration, pressure transients, signal
strength variation.
% Accepted The running tally of accepted up/down transmit burst sets.
Last Makeup Reason for last Makeup (signal reacquisition).
Makeup Status Current stage of Makeup routine. [Measurement = normal operation]