3-20
1010ENFM-3BSection 3
The 991TW insert sensors are designed to be used in pipes equipped with Thermowells. These are
spring-loaded, 1/4” diameter sensors with 1/2” NPT integral connection heads, available in several lengths
to accommodate a range of pipe sizes. Thermowells for new installations are available from Alloy Engi-
neering Company in Bridgeport, Connecticut.
3.2.1 WIRING TEMPERATURE SENSOR TO THE ANALOG INPUT MODULE
WARNING: Set flowmeter and instrumentation power to OFF when inserting or removing the
Analog Input Module, or when making connections to TB1, TB2, TB3, and TB4.
z Open the flowmeter top cover by releasing the cover latch.
z Place the power switch to the OFF position.
z Loosen the captive thumbscrew securing the Access Cover and remove Access Cover.
z Using a flat-blade screwdriver, remove four captive screws securing the I/O board.
z Carefully remove the I/O board and set it aside.
FUE1010EN Single Channel Temperature Sensor Inputs
TOP VIEW
SHORT TERMINALS 1 AND 4
SINCE T2 IS NOT BEING USED
- see note below.
TO SENSOR
NOTE: If this meter is using only one RTD tem-
perature sensor, short terminals 1 and
4 of TB2 with a jumper wire to com-
plete the dual RTD current loop.
Refer to paragraph 3.2.1 step 3
on next page.
Access Cover
Screw
FUE1010EN
Power Switich
Latch