6-4
1010ENFM-3B
Section 6
6.3 HOW TO SET UP FUE1010EN FOR FLOW TUBE OPERATION
The example in this section shows how to set up channel 1 of a 1010 dual channel system for use with
a 1011FT stainless steel Flow Tube. Note that Series 1010 dual-channel modelsallow flow tube opera-
tion with either measurement channel. In addition, all 1010 flow computers operate with both 1011FT
and 992DFT series of flow tubes. This example begins at the Channel Setup Menu. If necessary, review
paragraph 1.7.3 for instructions on how to select a Meter Type and a measurement channel.
6.3.1 OVERVIEW
Setting up the measurement channel for Flow Tube operation consists of the following tasks:
z Select a Meter Type (in this case, Dual Channel Flow).
z Select a channel to install the flow tube (e.g., Channel 1).
z Create a site setup to store the flow tube installation parameters.
z Enter liquid Parameters (optional, default = water@68°F).
z Complete the Flow Tube install procedure.
6.3.2 SETUP PROCEDURE
From the Channel Select screen, press to access the option list, then press to move
the cursor to [FlowTube] (if required).
To select Flow Tube press .This enables the Flow Tube menu. Press to access the
Channel Setup menu.
ENT
Channel 1 Clamp-on
Channel 2 >FlowTube
Reflexor
Siemens 2 Channel [1] Channel 1
Choose Channel 1 Flowmeter Type
Dual Channel Energy