2-87
1010ENFM-3B
Section 2
To mount transducers on a 996PSP Pipe Simulator
1. Using coupling compound, mount the transducers on the pipe simulator as shown on the previous
page. Slide each transducer until it presses against a pin-stop. Use the clamps to hold the transducer
in place.
2. Connect cables between each transducer and the computer connectors for the channel under test.
The Up and Down orientation is not important.
3. Press <MENU> to access the Installation Menu. Select [Meter Type] [Single, Dual, or Quad Channel]
depending on meter type. Select the meter channel (1,2,3, or 4) depending on the measurement
channel you intend to test. Select [Clamp-on] and then [Channel Setup].
4. Access [Channel/Path Setup] menu. Move highlight to [Create/Name Site]. Create a new Site Setup
(e.g., TEST1). You can now enter data without altering an existing Site Setup.
5. Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator chart below, en-
ter the pipe parameters corresponding to the transducer size under test.
2.11.11 USING THE 996PSP PIPE SIMULATOR
Using the proper coupling compound, mount the transducers on the pipe simulator as shown below.
Slide each transducer until it presses against a pin-stop. Use the clamping screws to hold the transducers
in place.
Clamping Screw
Xdcr
Clamping
Bracket
991 Transducers
Under test
Pipe Simulator
Transducer
Pinstops
Xdr
Clamping
Bracket
Clamping Screw
991 Transducers
Under Test
Transducer
Pinstops