2-18
1010ENFM-3B
Section 2
The flow computer will adjust its parameters to optimize performance based on your selections. The Ltn
menu cell shows the required spacing distance (in. or mm) between the upstream and downstream
transducers. Use the [Install Completed?] menu cell to inform the flow computer that you completed the
physical mounting of the transducers. You can define the empty pipe and zero flow values once the
transducers are operational.
NOTE: Before preceeding to mount transducers, it is recommended that Section 3 -Hardware
Installation Guide be reviewed. Refer to the Pick/Install Xdcr menu and menu struc-
ture shown below for menu cell descriptions and details.
Siemens 2 Channel [1] FLOW1
Scroll List and select desired Model
Pick/Install Xdcr
This menu cell
shows the actual
spacing distance
required between
the transducers.
Use this menu cell to
select the type of
transducer to be
installed.
Transducer Model 1011HG Hi Prec.
Transducer Size B2H
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 2
Spacing Method Track 1012TNH
Ltn Value (in) -1.244
Install Completed? Yes
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Actual Zero
This menu cell
establishes the
spacing between
transducers.
Changing the
Spacing Offset
alters the reported
Number index.
Once you select the
type, use this menu
cell to specify the
transducer type.
This menu cell
allows you to select
a mounting mode.
*Use Reflect Mode
whenever possible.
Use this menu cell to
inform the meter that
you have completed the
transducer mounting.
fter the transducers
are operating, you
can use this menu
cell to set the empty
pipe threshold.
After the transducers are
operating, you can use
this menu cell to manuall
set the zero flo
correction.
PICK/INSTALL XDCR MENU STRUCTURE
Pick/Install Xdcr Transducer Model 991 Universal
1011 Universal
1011 Hi Precision
Transducer Size
A1
A1 0A0A A1HA1H
A2 0 A2H
B1 1 A3H
B2 2A B1H
B3 2 B2H
C1 3A C1H
C2 3 C2H
C3 4A D1H
D1 4 D2H
D2 5A D4H
D3 5
E1
E2
E3
Xdcr Mount Mode Direct
Reflect
1011 Universal 991 Universal 1011H High Precision
Ú
Ö
Ö
Ö
Ö
Ú
Ú
Ú
Ú
Ú
Ú
Ú
Ú
(continued)