4-12
1010ENFM-3B
Section 4
When you are sure that the RTD Sensor is at 32°F (0°C), press   to recalibrate the RDT sensor.
To verify the calibrated reading, go to [Diagnostics/Liquid Data] to check the current RTD output. Make
sure that it coincides with the reading of the reference thermometer. Repeat for the other RTD, if neces-
sary.
ENT
4.7 THE CLOCK SET MENU
The Clock Set menu allows you to set the time and date. The meter uses its internal clock/calendar to
record the real-time when certain data and diagnostic events occur. In addition, the clock/calendar pro-
vides the Datalogger and Stripchart date and time stamps.
4.7.1 DATE
The Date command sets the month, day and year for the meter’s internal clock/calendar. Enter the date
using the [MM.DD.YY] format. Replace MM with two digits to indicate the month. Replace DD with two
digits to indicate the day. Replace YY with two digits to indicate the year. Type the period (.) separator
between each set of digits. For example, you would enter December 7, 1997 by typing 12.07.97. You
can use the <Right Arrow> to move the cursor back to a number if you need to re-type it. Press the
<CLR> key to remove the current setting.
To set the Date:
To enable numeric entry press
Use the numeric keys to type the date (MM.DD.YY).
To store the date press   . This moves the cursor to [Time].
4.7.2 TIME
The Time command sets the hours and minutes for the meter’s internal clock/calendar. Enter the time
using the [HH.MM.SS] format. Replace HH with 2 digits to indicate the hour (use 24-hr. clock format).
Replace MM with 2 digits to indicate the minutes. Replace SS to indicate the seconds. You also have to
type the period (.) separator between each set of digits. For example, you would enter ten minutes after
two o’clock in the afternoon by typing: 14.10.00 You can use the <Right Arrow> to move the cursor back
to a number if you 
need to re-type it. Press the <CLR> key to remove the current setting.
ENT
Calib @ deg F
=32.0
Notice:
All Siemens flowmeters include a real-time clock to provide a convenient date/time stamp for display screens and
datalogger reports. The operating system does not rely on the date and time-of-day for any flowmeter operation.
Therefore, the “Turn Of The Century” will have no effect on the proper functioning of any of our systems. Although
we limit the year to 2 digits in all our displays, 1010’s datalogger reports are Y2K complaint using an industry
standard algorithm to form a 4 digit year  field.
Clock Set Date xx.xx.xx (date entry)
CLOCK SET MENU STRUCTURE
Time  xx.xx (time entry)
Ö