Siemens SITRANS F - 2.3 The Application Data Menu

Siemens SITRANS F
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-11
1010ENFM-3B
Section 2
Use the Application Data menu to edit default settings to match your application. When you specify a
Liquid Class, the meter will adjust its operation to accommodate the liquid’s estimated sonic velocity,
viscosity and specific gravity. If necessary, you can edit each liquid class parameter individually to obtain
a closer match with the liquid. The default liquid is Water at 68
°F (20°C).
The [Temperature Range] menu cell allows you to specify the expected temperature range at the trans-
ducer mounting location. The default setting: [-40
0
F to 250
0
F], matches the standard 991 and 1011 ser-
ies of transducers but not the 1011H. For higher pipe temperatures, selecting the proper temperature
range allows the flow computer to recommend the appropriate transducers.
Energy Data
is provided
for liquids in
bold text
APPLICATION DATA MENU STRUCTURE
Application Data Liquid Class Select Liquid Water 13C/55F
Water 20C/68F
Water 50C/122F
Water 75C/167F
Water 100C/212F
Water 125C/257F
Water 150C/302F
Water 175C/347F
Water 200C/392F
Water 225C/437F
Water 250C/482F
10% Ethyl Glycol
20% Ethyl Glycol
30% Ethyl Glycol
40% Ethyl Glycol
50% Ethyl Glycol
27% CaCl Brine
Stm Condensate
Acetic Acid
Alcohol
Bromine
Carbon Tet
Chlorine
Diesel Fuel
Gasoline
Glycerine
Kerosene
MEK
Oil (SAE 20)
(continued)
Ö
Ú
Ö
Ö
Ú
ÚÚ
Use this menu to select
a class of liquids and
optionally edit the
following liquid
parameters:
x
Estimated Vs
x
Viscosity
x
Density
(Specific Gravity)
Siemens 2 Channel [1] Channel 1
Select Liquid Class from Liquid Table
Location Return
Service Heating
Temperature Range -40F to 250F
Pipe Configuration Fully Developed
Anomaly Diams 10
Liquid Class Water 20C/68F
Application Data
Use this menu cell
to define where the
flow transducers
will be installed.
Use this menu cell to enter the
number of pipe diameters
between the upstream
configuration and the
transducer installation.
Use this menu cell to select
the pipe configuration that
most accurately represents
the u
p
stream
p
i
p
e condition.
Use this menu cell
to specify the
temperature range
of the li
q
uid.
Use this menu cell
to specify the type
of energy service
being monitored by
the channel.