2-17
1010ENFM-3B
Section 2
2.4 THE PICK/INSTALL XDCR MENU
Use this menu after creating a new site setup in the Channel Setup menu, and defining the pipe param-
eters in the [Pipe Data] menu.
Based on pipe data (and optionally application data) entries, the Pick/Install Xdcr menu automatically
identifies the most suitable transducers for the application. It recommends the appropriate mounting
mode (direct or reflect) and lists the Spacer Bar or Mounting Track part number and spacing index.
Ideally, you will be able to use the primary recommendations. However, you can edit the menu entries as
required to accommodate different transducers or mounting configurations.
Application Data Pipe Configuration Fully Developed*
(Change to Upstream 1 Elbow
Piping) Dbl Elbow +
Dbl Elbow -
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Anomaly Diams xxxx (numeric entry)
Ö
Ú
Ö
Pipe Configuration Menu Structure
* Default
Ú
Ú
Ö
Pipe Configuration Option List Definitions
Fully Developed Fully Developed flow, as would be expected for very long straight pipe runs or
installation downstream of a flow condition.
1 Elbow Single 90 degree Elbow upstream of transducer installation.
Dbl Elbow + Double out-of-plane Elbows upstream of transducer installation.
Dble Elbow - Double in-plane Elbows upstream of transducer installation.
Valve To Be Determined.
Expander Pipe expansion upstream of transducer installation.
Reducer Pipe reduction upstream of transducer installation.
Norm Entry To Be Determined.
Header Inlet Header or pipe manifold upstream of transducer installation.
Intrusions To Be Determined.
Use this menu cell to select
the pipe configuration that
most accurately represents
the upstream pipe condition.
Siemens 2 Channel [1] Channel 1
Designate Pipe Shape Near Transducers
Liquid Class Water 20C/68F
Location
Return
Service Heating
Temperature Range -40F to 250F
Pipe Configuration Fully Developed
Anomaly Diams 1 Elbow
Dbl Elbow +
Dbl Elbow -
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Application Data
Use this menu cell to enter the
number of pipe diameters
between the upstream
configuration and the
transducer installation.