2-88
1010ENFM-3B
Section 2
NOTE: When using a 996PS-4 Simulator, you must access the [Application Data] menu, select
[Liquid Type] - [Other], then program the [Estimated Vs] for a value of 2100 m/s.
6. Access the [Pick/Install Xdcr] menu. Access the [Transducer Model] menu cell. Enter the Transducer
Model option list. Select [991 Universal].
7. Check the [Transducer Size] menu cell. If necessary, enter the Transducer Size option list and pick
the actual transducer under test.
8. Check the [Xdcr Mount Mode] menu cell. Adjust as necessary to match the simulator chart above.
9. Move the highlight down to [Install Completed?]. Access the option list. Move the cursor to [Install].
Press <ENT> to start the transducer install routine.
10. Check to make sure that the flow computer returns the Number Index listed in the chart above. Next,
review and verify the Diagnostics data items as described previously. If the computer, cables and
transducers are operating properly, then review all site conditions to locate the operating problem.
If a Pipe Simulator/Test- Block Test Fails:
1. Replace the transducer cables. If this allows you to complete the test as described above, then the
cables were defective.
2. If the cables are proven to be good, replace the transducers on the pipe simulator with a “known
good” set. If the system functions properly, then the original transducers under test are defective.
Please return any defective transducers to Siemens for repair.
3. If you replace the transducers and cables, and the system still fails to function correctly, then it is
likely that the computer has a malfunctioning circuit module. If you have spare modules you may try
module substitution to identify the defective module. Otherwise, please call the Technical Service
Department for further instructions.
Part Xdcr Pipe Pipe Wall Mount Letter Number
Number Size OD (in) Mat’l Thk (in) Mode Index Index*
996PSP-0 0 2.420 PVC 0.076 Direct A 6
996PSP-1 1 2.481 PVC 0.077 Direct A 9
996PSP-2 2 2.743 PVC 0.136 Direct A 6
996PSP-3 3 3.758 PVC 0.344 Direct A 1
996PSP-4 4 6.500 PVC 1.000 Direct** A 0
996PSP PIPE SIMULATOR CHART (English)
* The flow computer generates the Number Index.
** See note below.
* The flow computer generates the Number Index.
** See note below.
Part Xdcr Pipe Pipe Wall Mount Letter Number
Number Size OD (mm) Mat’l Thk (mm) Mode Index Index*
996PSP-0 0 61.47 PVC 1.93 Direct A 6
996PSP-1 1 63.02 PVC 1.96 Direct A 9
996PSP-2 2 69.67 PVC 3.45 Direct A 6
996PSP-3 3 95.45 PVC 8.74 Direct A 1
996PSP-4 4 165.1 PVC 25.4 Direct** A 0
996PSP PIPE SIMULATOR CHART (Metric)