2-72
1010ENFM-3B
Section 2
2.11.7 THE SITE SETUP DATA MENU
This menu provides data pertaining to transducer characteristics and operation. Some menu items are
for technical support interpretation only.
Reynolds # 310017
Siemens Dual Path SITE1
Current Reynolds number
Liquid Data
Site Setup Data
fx (drive) 30
N (burst length) 42
Ltn in -1.154
Vfmax GAL/MIN 1577.42
Vs max m/s 2165.41
Vs min m/s 936.62
Empty % 30
Samples/Cycle 24
HF 0.120
Siemens 2 Channel [1] SITE1
Current transit drive code
Reynolds # FUE1010EN corrects the current flow rate in accordance with this Reynolds num-
ber. This number is determined from pipe I.D., current flow rate and liquid kinematic
viscosity.
LIQUID DATA MENU ITEMS
fx (drive) Current Transmit drive code selected during Initial Makeup. The drive code con-
trols the sonic transmit signal.
N (burst length) Transmit burst duration selected during Initial Makeup. To change N count press
<Right Arrow>. At equal sign enter numeric value (1 to 9 only).
Ltn (in/mm) Spacing distance between the transducers. It will be in inches or millimeters, de-
pending on default units.
Vfmax The flow velocity (in selected units) corresponding to one whole cycle
offset between upstream and downstream receive signals.
Vs max m/s Maximum correctly calibrated Vs for current transducer spacing.
Vs min m/s Minimum correctly calibrated Vs for current transducer spacing.
Empty % Value of Empty Alarm Setting. The flowmeter will declare an empty status if signal
strength drops below this value.
Samples/Cycle Digital sampling rate.
Max Damping Maximum signal damping. Use to average digital data when an unstable condi-
tion occurs.
Min Damping Minimum signal damping. Use to average digital data when an unstable condition
occurs.
HF Flow registration correction factor.
SITE SETUP MENU ITEMS