6-6
1010ENFM-3B
Section 6
To move the highlight to [Install Flow Tube] press . Press to access the Install Flow
Tube menu.
To access the [Flow Tube] option list press the <Right Arrow>. Note that this list shows the
currently available flow tube sizes. The number is the size (e.g., 1). The letter represents the
material (e.g., S = Steel, P = PVC, T = Teflon, K = Kynar). In this example, we will install a size 1S
Flow Tube. This 1011FTP-1S (Stainless Steel) unit uses a 5/8” Swagelok® Tube Fitting.
To select size 1S press . This moves the highlight to [Install Completed?].
ENT
Siemens 2 Channel [1] FLOW1
Select FlowTube 1S
Install Completed? No
Empty FlowT
ube Set Chan Not Setup
Zero Flow Adjust Chan Not Setup
Pick Flow Tube size and Type
Install FlowTube
Siemens 2 Channel [1] FLOW1
Select FlowTube >1S
Install Completed? 1P
Empty FlowTube Set 1K
Zero Flow Adjust 1T
2S
2P
2K
2T
Pick Flow Tube size and Type
Install FlowTube
Note that the next menu down is Liquid Data. This menu allows you to enter Viscosity (in
centistokes) and the Density (specific gravity) data for the liquid to be measured. Data entry
in this menu is optional. The default values are those of Water at 68
°F (20°C). If you choose to
edit these parameters, you should be aware that the viscosity of the liquid is an important
factor. It governs the degree of Reynolds Number compensation that the flow computer
applies to the final rate output. Therefore, flow data errors could result if you enter inaccu-
rate data. Note that the Technical Service Group may be able to provide the viscosity data for
the liquid to be measured.