1-16
1010ENFM-3B
Section 1
The flowmeter is now ready to report flow. Press the <MENU> key. The meter will then ask you to
save the Site. Saving the Site is optional and will therefore be covered in paragraph 1.7. To abort this
item, press the <LEft Arrow>. This brings up a Flow Display Screen.
Once your site has successfully installed, you can press the <Up/Down Arrows> to scroll between
the available display screens.
Now that you are familiar with the FUE1010EN setup, take the time to read the information about the
Installation Menu in Section 2. Becoming acquainted with the various flowmeter setups in the Main
Menu will help further your understanding of the FUE1010EN.
Upon completion, the meter reports the measured liquid sonic velocity. Press the <Down Arrow>.
The flowmeter will then automatically perform an AutoZero operation, where the meter measures
any zero flow offset, which is used to correct the flow reading. This operation occurs only when the
transducers are mounted in the reflect configuration and is independent of whether or not the liquid
is flowing.
After the AutoZero operation is complete, the meter may ask you to register the transducer used in
this installation. It is advisable that you register your transducers, since the registration list is used to
condition the transducer recommendation.
Press <Right Arrow> and select [YES], or [NO] if you do not want to register this transducer and then,
press <ENT>.
Siemens 2 Channel [1] Unit#1
ENERGY KBTU/HR
TOTAL KBTU Td
500
0.0
0.63
- 5.23
0.74
Ve
1474
Siemens