2-22
1010ENFM-3B
Section 2
After transducers are properly mounted, the flow computer completes its Initial Makeup command and
the [Install Completed?] menu cell appears.
To access the [Install Completed?] option list, press the <Right Arrow>.
Press <Down Arrow> and scroll to the [InstallCompleted?] menu cell and then press <ENT>
(unless otherwise directed to do so by the Technical Service Department).
Press <ENT>. The [Install Completed?] menu cell will indicate [Yes] after the FUE1010EN is
successfully installed.
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? No
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Yes
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
NOTES
z Mount the transducers using the selected mode (direct or reflect). Please refer to the trans-
ducer mounting procedures detailed in Section 3.
z When using a mounting track or a spacer bar in reflect mode, locate the first transducer
at the Reference Index and the second transducer at the recommended Number Index.
z If you are mounting the transducers independently, you must use the recommended dis-
tance; Ltn to space the transducers.
z You must use the proper sonic coupling compound. See “Recommended Sonic Coupling
Compounds” in Section 5 for appropriate type and part number.
z The pipe must be completely filled with a liquid, which can be either flowing or at zero
flow.