2-30
1010ENFM-3B
Section 2
Siemens Dual Path SITE1
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Reflect
Spacing Offset Minimum Flow
Number Index 7
Spacing Method Track 1012TP
Ltn Value (in) 0.778
Install Completed? ZeroMatic
Empty Pipe Set Auto Zero
Zero Flow Adjust ZeroMatic
Zeroclr
ZeroMatic Active [6:.................:0]
Pick/Install Xdcr
When the inital makeup of ZeroMatic is complete the screen will return to the Pick/Install Xdcr menu
and automatically highlight [Operation Adjust], which is the next menu cell.
Channel/Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Stripchart Setup
Datalogger Setup
I/O Data Control
Diagnostic Data
Siemens Dual Path
Dual Path Flow
Conforms operation to user Preferences
Select the [ZeroMatic] menu cell by pressing or then press
ENT
Siemens Dual Path SITE1
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Reflect
Spacing Offset Minimum Flow
Number Index 7
Spacing Method Track 1012TP
Ltn Value (in) ReversaMatic
Install Completed? ZeroMatic
Empty Pipe Set Auto Zero
Zero Flow Adjust Actual Zero
ZeroClr
Conformans Indicated flow to Actual Zero
Pick/Install Xdcr