98 Copyright Silca 2004
IDEA Operating Manual - English
Fig. 125
Measurement X should be 7,60 mm (….inch), +/- 0,03 mm
Measurement Y (Y1+Y2) should be 20,40 mm (…inch) +/- 0,03 mm
If this is not the case, e.g. X = 7,50 and Y = 20,50 proceed as follows:
Choose “4 Calibrations” from the main menu and press ENTER.
Choose “1 Clamps” and press ENTER.
Choose “2) S100 – Reader” and press ENTER.
Digit the required corrections for axis X and axis Y (bit 1 = Y1 / bit 2 = Y2). Press ENTER.
Press ENTER to confirm the data.
4 - Calibrations
5 - Maintenance
6 - Options
1 – Clamps
2 – Cutters
3 – Adapters
Clamps
1: S1 - Rotating Jaw
2: S100 - Reader Jaw
Clamp: S100
Adjustments:
X= -10 Y1= -0,5 Y2= -0,5
To calibrate
(START)
Save
adjustment?
No=STOP Yes =ENTER
20,40
7,60