Commands and Queries
Programming Guide A-3
Full Description
This section explains the commands and queries in detail. The form of the command is
followed by the form of the associated query, which is followed by an explanation of
each.
ON Turn On Pump Laser
Turns on the pump laser.
NOTE: The shutter is not automatically opened when the ON command is issued.
The response to this command depends on whether or not the system is warmed up. Use
the READ:PCTWarmedup? query to determine the progress of the warm-up cycle.
When the response to this query reaches 100, the laser can be started. While the response
is 0 to 99, the ON command is simply ignored.
SERVO page A-6
SERVOTIME? page A-6
SHUTDOWN page A-6
SHUTter (n) page A-7
SHUTter? page A-7
TIMer:WATChdog (n) page A-7
WAVelength (nnn) page A-7
WAVelength? page A-7
WAVelength:min? page A-7
WAVelength:max? page A-7
*IDN? page A-7
*STB? page A-8
CONTrol:SLMIN? page A-9
CONTrol:SLMAX? page A-10
CONTrol:MTRMOV? page A-10
CONTrol:DSMPOSition? page A-10
If the response to
READ:PCTWarmedup? is...
The response to ON is...
0 to 99 The ON command is ignored.
100 The laser diode drive currents ramp to full power.