Objet260 User Guide 6 Operating and Maintaining the Objet260 3DPrinter
DOC-37004 Rev.E 6-67
IftheUVlevelmeasuredisnotacceptable,thewizardcalibratestheUVlamp
byadjustingthecurrentsuppliedtoit,andthentestsitagain.
WhenallUVtestshavefinished,theresultsaredisplayed,showingthe
conditionofthelampsaftercalibration(seeFigure6-100below).
Figure6-100ResultsandconditionofUVlampsaftercalibration
AfterexaminingtheUVcalibrationresults:
1. ClickNext.
2. RemovetheUVsensorandclosetheprintercover.Afterconfirmingthe
itemslistedinthewizardscreen,clickNext.
Figure6-101FinalUVCalibrationWizardscreen