Emergency Operation
WWaarrnniinngg:: The SwitchPoint Infinity 3 is designed with backup video. The purpose of
the backup video is to deliver a backup video signal to the display in the event the
system is unable to deliver the primary care video to the display.
The default primary and secondary care video source is the Endoscopic Camera. The
customer may work with Stryker during installation to have a different source connected
as the primary or secondary care video sources.
In the event that the primary care video is lost to the display(s), and manual rerouting or
use of preset fails to restore the video, the user should switch the system to backup
video mode, using the following steps (in any order):
1. Manually switch the video mode of Monitor 1 to Analog RGB. This is done on
Monitor 1.
2. Manually switch the video mode of Monitor 2 to Composite or S-Video. This is
done on Monitor 2.
Performing these steps will deliver the primary care video to Monitor 1 and a secondary
video (video source may be the same as that of Monitor 1) to Monitor 2.
Promptly contact Stryker Communications for assistance in permanently resolving the
issue.