EEmmiissssiioonnss TTeesstt CCoommpplliiaannccee EElleeccttrroommaaggnneettiicc EEnnvviirroonnmmeenntt --
GGuuiiddaannccee
RF emissions
CISPR 11
Group 1 The SwitchPoint Infinity® 3 Control
System uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A The SwitchPoint Infinity® 3 Control
System is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/ flicker
emissions
IEC 61000-3-3
Not applicable
GGuuiiddaannccee aanndd MMaannuuffaaccttuurreerrss DDeeccllaarraattiioonn -- EElleeccttrroommaaggnneettiicc IImmmmuunniittyy
The SwitchPoint Infinity® 3 Control System is intended for use in the electromagnetic
environment specified below. The customer or the user of the SwitchPoint Infinity® 3
Control System should assure that it is used in such an environment.
IImmmmuunniittyy
tteesstt
IIEECC 6600660011
tteesstt lleevveell
CCoommpplliiaannccee
lleevveell
EElleeccttrroommaaggnneettiicc eennvviirroonnmmeenntt --
gguuiiddaannccee
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
+/- 6 kV
contact
+/- 8 kV air
+/- 6 kV
contact
+/- 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical
fast
transient/
burst
IEC61000-
4-4
+/- 2 kV for
power supply
lines
+/- 1 kV for
input/output
lines
+/- 2 kV for
power supply
lines
+/- 1 kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-
4-5
+/- 1 kV line
(s) to line(s)
+/- 2 kV line
(s) to earth
+/- 1 kV
differential
mode
+/- 2 kV
common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.