92
Voltage
dips, short
interruptions
and voltage
variations
on power
supply input
lines.
IEC 61000-
4-11
<5% UT
(>95% dip in
UT) for 0,5
cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
(30% dip in
UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 sec
5% UT
(95% dip in
UT) for 0,5
cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
(30% dip in
UT)
for 25 cycles
5% UT
(95% dip in
UT)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
SwitchPoint Infinity® 3 Control System
requires continued operation during
power mains interruptions, it is
recommended that the SwitchPoint
Infinity® 3 Control System be powered
from an uninterruptible power supply or
a battery.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
4-8
3 A/m 3 A/m Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NNoottee:: NNoottee UUTT iiss tthhee aa..cc.. mmaaiinnss vvoollttaaggee pprriioorr ttoo aapppplliiccaattiioonn ooff tthhee tteesstt lleevveell..