OPERATION OF THE MACHINE
MA6 / MA8 - 1. Operation of the Machine - Rev.05 07-06
1
34
1.5.6 Change Mask
For example: manual
Display message Following operator action (keystroke)
"ready for Load" CHANGE MASK
"Change Mask -
press ENTER to toggle
mask vacuum: OFF"
pull out mask holder and flip it 180°
deposit it on the tray left
place the mask against the stop pins
press ENTER
"Change Mask -
press ENTER to toggle
mask vacuum: ON"
activate the mechanical mask clamping
by pressing the leaf spring
flip mask holder 180° and slide it back
into the machine (up to the end and
preverably to the right position)
press CHANGE MASK
"ready for Load"