A Trigger 
SLOPE 
A Trigger Mode 
Appendix 
C: 
Performance Verification 
Positive 
(button out) 
NORM 
L 
r 
r 
i? 
i? 
r 
i? 
i? 
r 
E? 
L? 
r? 
i? 
E? 
r 
r 
E? 
2 
E? 
i2 
i2 
c 
i? 
i? 
i? 
z 
2232 
User Manual 
A-49 
-- 
- 
- 
- 
- 
- 
-- 
- 
- 
-- 
- 
- 
- 
- 
- 
- 
b. 
Set the generator to produce a 
50 
kHz, 
6.4 
division display. 
c.  CHECK 
- 
Display is triggered along the entire positive slope of the 
waveform as the A Trigger 
LEVEL 
control is rotated. 
d.  CHECK 
- 
Display is not triggered (no trace) at either extreme of 
rotation. 
e. 
Set the A Trigger 
SLOPE 
button to Negative (button in). 
f.  CHECK 
- 
Display is triggered along the entire negative slope of the 
waveform as the A Trigger 
LEVEL 
control is rotated. 
g.  CHECK 
- 
Display is not triggered (no trace) at either extreme of 
rotation. 
Step 
6: 
Check Single Sweep Operation 
a. 
Adjust the A Trigger 
LEVEL 
control to obtain a stable display. 
b. 
Set: 
CH 
1 
AC-GND-DC 
A Trigger 
SLOPE 
A 
& 
6 
SOURCE 
A COUPL 
A 
SECIDIV 
GND 
Positive 
(button out) 
CH 
1 
NORM 
20 
ms 
c. 
Press in the 
SGL SWP 
button. The 
READY 
LED should illuminate 
and remain on. 
d.  Set the 
CH 
1 
AC-GND-DC 
switch to DC. 
NOTE 
The 
A INTENSITY 
control may require adjustment to observe the 
single-sweep trace. 
e.  CHECK 
- 
READY 
LED goes out and a single sweep occurs. 
f. 
Press in the 
SGL SWP 
button several times. 
g.  CHECK 
- 
Single-sweep trace occurs, and the 
READY 
LED illumi- 
nates briefly every time the 
SGL SWP 
button is pressed in and 
released. 
h.  Disconnect the test equipment from the instrument.