-p 
- 
- 
- 
- 
- 
- 
- 
- 
- 
Displaying Signals 
Initial Data 
Direction of "Roll" 
C------ 
Current Data 
Figure 3-7:  Roll Mode 
Capturing 
Random 
Events 
Infrequent or random events can be "captured" by using the single-sweep 
trigger mode in combination with either the Scan or the Roll mode. These 
two modes are called "Scan-roll-scan" and "Triggered Roll" respectively. 
These two modes are only available at sweep speeds of 
0.1 
ms and slower. 
Triggered Roll: 
Step 
1: 
Set the 
A 
Trigger Mode to 
P-P AUTO. 
Step 
2: 
Press the Acquisition 
TRIG POS 
button to select the position on 
the screen where the trigger event ("T") will be displayed. 
Step 3: 
Press the Setup 
ACQ 
button to call up the Acquisition Menu. 
Step 
4: 
Select 
Roll 
and press the 
ACQ 
menu button again to exit the 
menu. 
Step 
5: 
Press the 
SGL SWP 
button. 
Note that the trace moves across the screen continuously from right to left. 
Also, the 
READY 
light is on indicating the oscilloscope is waiting for a trigger 
signal. (If the 
READY 
light is not on, press the 
SGL SWP 
button again.) 
When the trigger event occurs, the acquisition continues across the screen 
from right to left until it reaches the trigger point indicator. The oscilloscope 
then records the event coincident with the trigger, completes the record, and 
freezes the display. 
Step 
1 
: 
Set the 
A 
Trigger Mode to 
PP AUTO. 
Step 
2:  Press the Acquisition 
TRIG POS 
button to select the position on 
the screen where the trigger event 
("T") will be displayed. 
0 
Step 
3: 
Press the Setup 
ACQ 
button to call up the Acquisition Menu. 
In Detail