Appendix 
C: 
Performance Verification 
Adjust the A Trigger 
LEVEL 
control for a stable display and center 
the waveform on the screen. 
Set the 
CH 
1 
VOLTSIDIV 
switch to 10 mV for a 10 division display. 
Vertically position the top of the waveform display on the center 
horizontal graticule line. 
Set the A Trigger 
SLOPE 
switch to Negative (button in). 
Rotate the A Trigger 
LEVEL 
control clockwise until the triggering of 
the waveform display becomes unstable. 
CHECK 
- 
That the trigger readout is between 
92 
mV and 108 mV. 
Repeat procedure for 
CH 
2 
using the 
CH 
2 
controls. 
Disconnect the test equipment from the instrument. 
External 2-Axis, Probe Adjust, External Clock, and 
X-Y 
Plotter 
Checks 
Equipment Required (see Table A-26): 
Leveled Sine Wave Generator 
Two 50 
C2 
BNC Cables 
BNC T-Connector 
50 
C2 
BNC Termination 
Pulse Generator  BNC male-to-tip plug 
Digital Voltmeter 
1 
OX  Probe (provided with instrument) 
Initial Control Settings: 
Vertical (Both Channels) 
CH 
1 
POSITION 
MODE 
x-Y 
BW LIMIT 
CH 
1 
VOLTSIDIV 
CH 
1 
VOLTSIDIV 
Variable 
CH 
1 
AC-GND-DC 
Midrange 
CH 
1 
Off 
(button out) 
Off 
(button out) 
1 
v 
CAL 
detent 
DC 
Horizontal 
POSITION 
MODE 
A 
SECIDIV 
SECIDIV 
Variable 
XI0 
Magnifier 
Midrange 
A 
20 
ps 
CAL 
detent 
Off 
(knob in) 
A Trigger 
VAR 
HOLDOFF 
Mode 
SLOPE 
LEVEL 
A 
& 
B SOURCE 
A COUPL 
NORM 
P-P AUTO 
Positive 
(button out) 
Midrange 
Vertical 
MODE 
NORM