Appendix 
C: 
Performance Verification 
instrument that is operating outside this temperature range. All tolerances 
specified are for the instrument only and do not include test-equipment 
error. 
Test Equipment 
Required 
Table A-7 lists all the test equipment required to do the Performance Verifica- 
tion in this appendix. Also listed is the minimum specifications for the test 
equipment. All equipment used must meet or exceed its minimum specifica- 
tions. 
When equipment other than that recommended is used, control settings of 
the test setup might need to be altered. If the exact item of equipment given 
as an example in Table A-7 is not available, check the Minimum Specification 
column to determine if any other available test equipment might suffice to 
perform the check or adjustment. 
Operating instructions for test equipment are not given in this procedure. 
If 
more operating information is required, refer to the appropriate test equip- 
ment instruction manual. 
Table 
A-7: 
Test Equipment Required 
Item and  Minimum  Example of Suitable 
Description  Specification  Purpose  Test Equipment 
Calibration  Standard-amplitude signal levels: 5 mV to 50 V.  Signal source for  TEKTRONIX PG 
506A 
Generator  Accuracy +-0.3%.  gain and transient  Calibration Genera- 
High-amplitude signal levels: 1 V to 60 V.  response.  tor.' 
Repetition rate: 1 kHz. 
Fast-rise signal level: 1 
V. 
Repetition rate: 1 MHz. 
Rise time: 1 ns or less. Flatness: 
?2%. 
Leveled  Frequency: 250 kHz to above 100 MHz. 
Sine Wave  Output amplitude: variable from 10 
mV to 
5V 
p-p. 
Generator  Output impedance: 50 
Q. 
Reference frequency: 50 kHz. 
Amplitude accuracy: constant within 3% of refer- 
ence frequency as output frequency changes. 
Vertical, horizon-  TEKTRONIX SG 503 
tal, and triggering 
Leveled Sine Wave 
checks and ad-  Generator.' 
justments. Display 
adjustments and 
Z-Axis check. 
Time-Mark 
Marker outputs: 10 ns to 0.5 s. 
Generator 
Marker accuracy: 
+ 
0.1 
%. 
Trigger output: 1 ms to 0.1 ms, time-coincident 
with markers. 
Horizontal checks  TEKTRONIX TG 501 
and adjustments.  Time-Mark Genera- 
Display adjust-  tor.' 
ment. 
Low-  Range: 1 kHz to 500 kHz.  Low-frequency  TEKTRONIX SG 502 
4 
~scillator.~ 
!!? 
Frequency  Output amplitude: 300 mV.  trigger checks. 
Generator  Output impedance: 600 W. 
4 
Reference frequency: constant within 0.3 
dB 
of 
Lr. 
reference frequency as output frequency 
- 
changes. 
5 
pp 
- 
- 
- 
- 
- 
- 
- 
- 
4 
Pulse  Repetition rate: 1 kHz. Output amplitude: 5 V.  External clock and  TEKTRONIX PG 501 
tr. 
Generator  storage checks.  Pulse Generator.' 
.-.I 
* 
*r 
A-26 
Appendices 
2 
b