Appendix 
C: 
Performance Verification 
i. 
CHECK 
- 
Trace shift of 0.5 division or less. 
j. 
Press in the 
SAVEICONT 
button to select 
CONT 
k.  Position the trace 0.5 division above the bottom horizontal graticule 
line using the 
CH 
2 
POSITION 
control. 
I. 
Press in the 
SAVEICONT 
button to select 
SAVE. 
m.  CHECK 
- 
Trace shift of 0.5 division or less. 
n.  Press in the 
SAVEICONT 
button to select 
CONT 
o.  Set the Vertical 
MODE 
switch to 
CH 
1 
p.  Repeat steps g through m for 
CH 1 
trace. 
Step 
6: 
Check Bandwidth 
a.  Set: 
VOLTSIDIV 
(both) 
2 
mV 
AC-GND-DC 
(both) 
DC 
A 
SECIDIV 
0.2  ms 
BW  LIMIT 
Off 
(button out) 
STOREINON-STORE 
NON-STORE 
(button 
out) 
b.  Connect the leveled sine wave generator output via a 50 
fl 
precision 
cable and a 50 
fl 
termination to the 
CH 
1 
OR 
X 
input connector. 
c.  Set the generator to produce a 50 kHz, six division display. 
d.  CHECK 
- 
Display amplitude is 
4.2 
divisions or greater as the gener- 
ator output frequency is increased up to the value shown in Table 
A-1 0 for the corresponding 
VOLTSIDIV 
switch setting. 
Table A-1 
0: 
Settings for Bandwidth Checks 
VOLTSIDIV 
Generator 
Switch Setting  Output Frequency 
2 
mV 
80 MHz 
5 
mV to 0.5 V  100 MHz 
e.  Repeat parts c and d for all indicated 
CH 1 VOLTSIDIV 
switch 
settings, up to the output-voltage upper limit of the sine wave gener- 
ator being used. 
f. 
Move the cable from the 
CH 1 
OR 
X 
input connector to the 
CH 
2 
OR 
Y 
input connector. 
g.  Set the Vertical 
MODE 
switch to 
CH 
2. 
h.  Repeat parts c and d for all indicated 
CH 
2 
VOLTSIDIV 
switch 
settings, up to the output-voltage upper limit of the sine wave gener- 
ator being used.