Operation
Diagnostics Menu
3-78 Model 48i Instruction Manual Thermo Electron Corporation
Interface Board Voltages The Interface Board screen (read only) is used to display the current voltage
readings on the interface board.
In the Main Menu, choose Diagnostics > Voltages > Interface Board
Voltages.
Press to return to the Voltages menu or to return to the
Run screen.
I/O Board Voltages The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only accessible if the I/O
expansion board is installed.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.3 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
INTERFACE BOARD VOLTAGES:
> 3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-15.0 SUPPLY -15.0 V
18.0 IR SUPPLY 18.0 V
18.0 MOT SUPPLY 18.0
ª
RANGE AVG DIAGS ALARM
BIAS SUPPLY -110.0 V