Operation
Calibration Factors Menu
Thermo Electron Corporation Model 48i Instruction Manual 3-19
Calibration Factors
Menu
Calibration factors are used to correct the CO concentration readings that
the instrument generates using its own internal calibration data. The
Calibration Factors menu displays the calibration factors. The screens below
show the calibration factors menu in single mode and dual/autorange
modes.
Normally, the instrument is calibrated automatically using the Calibration
menu described in “Calibration Menu” later in this chapter. However, the
instrument can also be calibrated manually using the Calibration Factors
menu.
To manually calibrate the instrument, see “CO Backgrounds” and “CO
Span Coefficients” below for more information.
In the Main Menu, choose Calibration Factors.
Use and to move the cursor up and down.
Press to accept the choice.
Press to return to the Main Menu or to return to the
Run screen.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
AVERAGING TIME:
CURRENTLY: 30 SEC
SET TO: 10 SEC ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM