C-Link Protocol Commands
Measurement Configuration
Thermo Electron Corporation Model 48i Instruction Manual B-27
Measurement
Configuration
range
high range
low range
These commands report CO range in single range mode, or the high and
low ranges in dual or auto range mode. If the mode is incorrect, the
instrument responds with “can’t, wrong settings”. The example below
reports that the CO full-scale range to 50 ppm, according to Table B-6.
Send: range
Receive: range 6: 5000E-2 ppm
set range Selection
set high range Selection
set low range Selection
These commands select the CO full-scale ranges, according to Table B-6.
The example below sets the CO full-scale range to 50 ppm.
Send: set range 5
Receive: set range 5 ok
custom range
Table B-6. Standard Ranges
Selection ppm mg/m
3
011
122
255
31010
42020
55050
6 100 100
7 200 200
8 500 500
9 1,000 1,000
10 2,000 2,000
11 5,000 5,000
12 10,000 10,000
13 C1 C1
14 C2 C2
15 C3 C3