C-Link Protocol Commands
Measurement Configuration
B-28 Model 48i Instruction Manual Thermo Electron Corporation
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The example below reports that custom range 1 is defined to 5.50 ppm.
Send: custom 1
Receive: custom 1 550E-2 ppm
set custom range range value
set custom 1 range value
set custom 2 range value
set custom 3 range value
These commands are used to set the maximum concentration for any of the
three custom ranges 1, 2, or 3 to range value, where value is a floating-point
number representing concentration in ppm or mg/m
3
. The example below
sets the custom 1 range to 55.5 ppm.
Send: set custom 1 range 55.5
Receive: set custom 1 range 55.5 ok
range mode
This command reports the current range mode.
Send: range mode
Receive: range mode single
set range mode mode
This command sets the current range mode to single, dual, or auto. The
example below sets the range mode to single.
Send: set range mode single
Receive: set range mode single ok
gas mode
This command reports the current mode of sample, zero, or span. The
example below reports that the gas mode is sample.
Send: gas mode
Receive: gas mode sample
set sample
This command sets the zero/span valves to the sample mode. The example
below sets the instrument to sample mode, that is, the instrument is reading
the sample gas.
Send: set sample
Receive: set sample ok