EasyManua.ls Logo

Thermo 48i - References; HI and lo Multi-Point Calibration

Thermo 48i
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
References
Thermo Electron Corporation Model 48i Instruction Manual 4-7
2. Periodically challenging the instrument with a CO level of
approximately 80% of the URL. The 80% URL level may be obtained
by dilution of a higher level of CO using a system similar to that of
Figure 4-1, or by using a low level cylinder of CO containing CO in air
at a concentration of approximately 80% of the URL. In either case the
cylinder of CO should be checked against an SRM or CRM. It should
also be true for a cylinder of low level CO.
The Quality Assurance Handbook
1
should be referred to for the cylinder
checking procedure.
Record the instrument’s response in % of scale as A
80
. Compute the
span error from the following equation:
Where:
Z = Recorder response obtained at the last calibration for zero air, % scale.
[CO] = Span concentration
3. Latest copy of the Quality Assurance Handbook for Air Pollution
Measurement Systems
1
should be consulted to determine the level of
acceptance of zero and span errors.
For detailed guidance in setting up a quality assurance program, refer to
the Code of Federal Regulations
3
and the EPA Handbook on Quality
Assurance
1
.
References 1. QUALITY ASSURANCE HANDBOOK FOR AIR POLLUTION
MEASUREMENT SYSTEMS, Volume II - Ambient Air Specific
Methods EPA 600/4-77-027a, May 1977 and 40 CFR 50, Appendix C.
HI and LO Multi-Point
Calibration
The instrument can also be calibrated using a dual three-point calibration.
The calibration points are divided into a “low” range and a “high” range
consisting of three points each
•Cal Point 1
•Cal Point 2
Cal Point 3
Span Error, % = ([A
80
- Z)URL/100] - [CO]) X 100/[CO]

Table of Contents

Related product manuals