C-Link Protocol Commands
Record Layout Definition
B-40 Model 48i Instruction Manual Thermo Electron Corporation
Note if individual relays have been assigned different logic then the response
would be a 4-digit hexadecimal string with the least significant byte (LSB)
being relay no 1.
For example:
Receive: relay stat 0x0001 (indicates relay no 1 is set to normally
open logic, all others are normally closed)
Receive: relay stat 0x0005 (indicates relay no 1 and 3 are set to be
normally open logic, all others are normally closed)
set relay open
set relay open value
set relay closed
set relay closed value
These commands set the relay logic to normally open or closed for relay
number value, where value is the relay between 1 and 16. The example
below sets the relay no 1 logic to normally open.
Note if the command is sent without an appended relay number then all the
relays are assigned the set logic of normally open/closed.
Send: set relay open 1
Receive: set relay open 1 ok
Record Layout
Definition
The Erec, Lrec Srec layouts contain the following
A format specifier for parsing ASCII responses
A format specifier for parsing binary responses,
In addition to these the Erec Layout contains
A format specifier for producing the front-panel displays.
In operation, values are read in using either the ASCII or binary format
specifiers and converted to uniform internal representations (32-bit floats or
32-bit integers). These values are converted into text for display on the
screen using the format specifier for the front-panel display. Normally, the
specifier used to parse a particular datum from the input stream will be
strongly related to the specifier used to display it (e.g., all of the floating
point inputs will be displayed with an 'f' output specifier, and all of the
integer inputs will be displayed with a 'd' specifier).