C-Link Protocol Commands
Alarms
Thermo Electron Corporation Model 48i Instruction Manual B-11
Figure B-1. Flags
Alarms alarm agc intensity min
alarm agc intensity max
These commands report the agc intensity alarm minimum and maximum
value current settings. The example below reports that the agc intensity
alarm minimum value is 20.
Send: alarm agc intensity min
Receive: alarm agc intensity min 20
set alarm agc intensity min value
set alarm agc intensity max value
These commands set the agc intensity alarm minimum and maximum
values to value, where value is a floating-point number representing agc
intensity alarm limits. The example below sets the agc intensity alarm
maximum value to 20.
Send: set alarm agc intensity max 20
Receive: set alarm agc intensity max 20 ok
alarm bias voltage min
alarm bias voltage max
µ