C-Link Protocol Commands
Alarms
B-12 Model 48i Instruction Manual Thermo Electron Corporation
These commands report the bias voltage alarm minimum and maximum
value current settings. The example below reports that the bias voltage alarm
minimum value is 20.
Send: alarm bias voltage min
Receive: alarm bias voltage min 20
set alarm bias voltage min value
set alarm bias voltage max value
These commands set the bias voltage alarm minimum and maximum values
to value, where value is a floating-point number representing bias voltage
alarm limits. The example below sets the bias voltage alarm maximum value
to 20.
Send: set alarm bias voltage max 20
Receive: set alarm bias voltage max 20 ok
alarm chamber temp min
alarm chamber temp max
These commands report the chamber temperature alarm minimum and
maximum value current settings. The example below reports that the
chamber temperature alarm minimum value is 35.0 °C.
Send: alarm chamber temp min
Receive: alarm chamber temp min 35.0 deg C
set alarm chamber temp min value
set alarm chamber temp max value
These commands set the chamber temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing chamber temperature alarm limits in degrees C. The example
below sets the chamber temperature alarm maximum value to 55.0 °C.
Send: set alarm chamber temp max 55.0
Receive: set alarm chamber temp max 55.0 ok
alarm conc min
alarm conc max
These commands report the CO concentration alarm minimum and
maximum values current setting. The example below reports that the CO
concentration minimum is 5.2 ppm.
Send: alarm conc min
Receive: alarm conc min 5.2 ppm
set alarm conc min value
set alarm conc max value