C-Link Protocol Commands
Calibration
B-22 Model 48i Instruction Manual Thermo Electron Corporation
This command reports the currently set time interval in seconds for
streaming data.
Send: stream per
Receive: stream per 10
set stream per number value
number value = | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to number value in seconds. The example below sets the number
value to 10 seconds.
Send: set stream per 10
Receive: set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to Table B-5.
Send: stream time
Receive: stream time 0
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to Table B-5. The example below
attaches a time stamp to streaming data.
Send: set stream time 0
Receive: set stream time 0 ok
Calibration set cal co bkg
This command will auto-calibrate the CO background. The example below
shows a successful auto-calibration of the CO background.
Send: set cal co bkg
Receive: set cal co bkg ok
set cal co coef
Table B-5. Stream Time Values
Valu e Stream Time
00 Attaches time stamp to streaming data string
01 Disables time stamp to streaming data string