C-Link Protocol Commands
Datalogging
B-16 Model 48i Instruction Manual Thermo Electron Corporation
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send: diag volt iob
Receive: diag volt iob 24.1 4.9 3.2 -3.2
Datalogging clr records
This command will clear all long and short records that have been saved.
Send: clear records
Receive: clear records ok
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example below clears short records.
Send: set clr srecs
Receive: set clr srecs ok
list din
list dout
These commands report the current selection for the digital outputs in the
format. Output no Index number variable name active state. The active
state for digital outputs is open or closed. The active state for digital inputs
is high or low.
Send: list dout
Receive: list dout
output index variable state
1 22 I/O BD COMM closed
2 2 LOCAL/REMOTE open
3 4 UNITS open
4 19 BIAS VOLTAGE closed
7 7 SAMPLE MODE open
8 8 GEN MODE open
list lrec
list srec
list stream
list sp
These commands report the list of current selections for long record logging
data, short record logging data, streaming data output, or the scratch pad
(sp) list. The example below shows the list for streaming data output.