Operation
Diagnostics Menu
Thermo Electron Corporation Model 48i Instruction Manual 3-77
Voltages The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board item is only
displayed if the I/O expansion board is installed.
• In the Main Menu, choose Diagnostics > Voltages.
• Use and to move the cursor up and down.
• Press to select a choice.
• Press to return to the Diagnostics menu or to return to
the Run screen.
Motherboard Voltages The Motherboard screen (read only) is used to display the current voltage
readings on the motherboard.
• In the Main Menu, choose Diagnostics > Voltages > Motherboard
Voltages.
• Press to return to the Voltages menu or to return to the
Run screen.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
RANGE AVG DIAGS ALARM