Operation
Instrument Controls Menu
Thermo Electron Corporation Model 48i Instruction Manual 3-29
Zero/Span Ratio The Zero/Span Ratio screen is used to adjust the ratio of zeros to spans. For
example, if this value is set to 1, a span check will follow every zero check. If
this value is set to 3, there will be two zero checks between each zero/span
check. This value may be set from 1 to 10, with 1 as default.
In the Main Menu, choose Calibration > Zero/Span Check >
Zero/Span Ratio.
Use and to increment and decrement the numeric value.
•Press to save the ratio value.
Press to return to the Zero/Span Check menu or to
return to the Run screen.
Instrument Controls
Menu
The Instrument Controls menu contains a number of items. The software
controls listed in this menu enable control of the listed instrument
functions.
In the Main Menu, choose Instrument Controls.
Use and to move the cursor up and down.
Press to accept the choice.
Press to return to the Main Menu or to return to the Run
screen.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
ZERO/SPAN RATIO:
CURRENTLY: 1 :1
SET TO: 3 :1 ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM