Operation
Service Menu
Thermo Electron Corporation Model 48i Instruction Manual 3-113
Preamp Board Calibration The Pre-amp Board Calibration screen is used to adjust the preamp board
calibration parameters.
In the Main Menu, choose Service > Preamp Calibration.
Use and to increment or decrement the numeric value.
Press to save the new counts to hardware.
Press to return to the Service menu or to return to the
Run screen.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMX
COEFFICIENTS:
1: 1.000000
2: 1.000000
3: 1.000000
RECALCULATE?
RANGE AVG DIAGS ALARM
WARNING This adjustment should only be performed by an instrument
service technician.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
PREAMP BOARD CAL:
SAMPLE: 127807
REFERENCE: 148256
SET TO: 128 ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM