Operation
Alarms Menu
3-98 Model 48i Instruction Manual Thermo Electron Corporation
Max Zero and Span Offset The Max Zero Check Offset screen is used to set the maximum zero check
offset. The maximum zero and span offset screens function the same way.
In the Main Menu, choose Alarms > Zero or Span Check > Max Offset.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
Press to return to the Zero or Span Check screen or to
return to the Run screen.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
ZERO CHECK:
ALARM: OK
RESPONSE: 0.108
>MAX OFFSET: 0.200
RANGE AVG DIAGS ALARM
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
MAX ZERO CHECK OFFSET:
CURRENTLY: 0.200
SET TO: 0.500 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM