Calibration
HI and LO Multi-Point Calibration
4-10 Model 48i Instruction Manual Thermo Electron Corporation
14. Press to back up a step in the Service Mode menu.
15. Select Coefficients and press .
The instrument will automatically calculate the new coefficients and
save the new parameters.
16. Repeat Steps 1 through 13 for Hi Multi-Point Cal.