System Description
Software
8-6 Model 48i Instruction Manual Thermo Electron Corporation
The analog output ranges are user selectable via software. The analog
outputs are defaulted based on the measurement range. The defaults are
calculated by dividing the data values by the full-scale range for each of the
three parameters and then multiplying each result by the user-selected
output range. Negative concentrations can be represented as long as they are
within -5% of full-scale. The zero and span values may be set by the user to
any desired value.