5.
Click Done.
The run does not proceed under any of the following conditions:
The waste bin is not in position
The sample plate is not in position
The OpenArray
Plates are not in position
There are more OpenArray
Plates on the instrument deck than are defined in the experiment
setup
The instrument door is open
The instrument will proceed to load the next OpenArray
Plate.
6.
Repeat step 1 to step 5 for each OpenArray
Plate to be loaded.
After all of the OpenArray
Plates have been loaded, the Deck screen displays Run completed
successfully. Empty the waste bin before performing another run.
If the run was started with a TPF or SPF file, a loaded TPF or SPF file is generated for each OpenArray
Plate. The loaded TPF or SPF file corresponds to the original TPF or SPF file that was imported for
the run. The files are exported to the folder that was designed in the Preferences (see “Set the folder
defaults” on page 15).
If the run was started without a TPF or SPF file, a loaded TPF or SPF file is not generated.
Chapter 9 Prepare the OpenArray
Plate for PCR
Remove the OpenArray
Plate from the instrument
9
60
QuantStudio
12K Flex OpenArray
AccuFill
System User Guide