88
BEFORE REQUESTING SERVICE
BEFORE REQUESTING SERVICE
TROUBLESHOOTING
Messages displayed during operation
Troubleshooting
When an error is encountered, review the Check List below.
If, after following the instructions below, you still cannot restore the instrument to
a normal condition or if the problem does not fall into any of the categories below,
contact your dealer or TOPCON (see the back cover).
Check List
"CHECK COVER" The lamp house cover is off.
"CHARGE ERROR
POWER OFF"
Overcharge, undercharge; or Fuse (F4) is burnt.
"CAMERA ERROR" The CCD camera has malfunctioned.
"CHECK CF" The CompactFlash
®
card has malfunctioned.
"CHECK USB" The photographed image is not transferred to the computer.
"CHECK DATA" The input of the optional numerals and ID data is not completed
normally.
"USB NOT CONNECT" The USB cable is not connected properly.
"CARD FULL" The capacity of the CompactFlash
®
card is full.
"NO CARD" The CompactFlash
®
card is not inserted.
"PC NOT READY" The computer is not in the proper condition to enable photography.
"CHECK MODE DIAL" The mode selector knob is changed to others except the PC mode
when the STORAGE mode is in use.
"STORAGE" The STORAGE mode is set in the PC mode.
"NO IMAGE" There is no playback image in the playback mode.
"PROTECTED FILE" The image to be deleted is protected.
"DPOF!" The files beyond 999 files are specified for printing.
"INVALID FILE" The file is invalid or damaged.
WARNING
To avoid electric shock, do not attempt disassembling, re-
building and/or repairs on your own. Ask your dealer for
repairs.
WARNING
Do not remove the covers from the main unit, chinrest unit
or power supply unit except for the lamp house cover.
You may receive an electric shock.
Trouble Condition Check Page
Color video monitor
does not work.
Power cable is off the receptacle/
instrument.
Connect power cable.
21
Fuse is burnt. Change fuse. 96
Power save function is on (flickering
power lamp).
Press Photography switch and cancel
power save function.
41